"FDA-cleared AI" is the most convenient trust claim in health tech marketing. It borrows the credibility of an agency that approves cancer drugs, and it works because almost nobody checks what the clearance actually says. The number behind the claim is now big enough to be a headline in itself: the FDA's public list of AI-enabled medical devices passed 1,600 entries in its September 2026 update (FDA). The same registry also contains a second number that gets quoted far less. As of August 2026, a peer-reviewed census of the cleared devices found that exactly three of them had been evaluated in registered studies for what happens to patients: mortality, morbidity, or readmissions (PLOS Digital Health).
One thousand six hundred clearances. Three outcome trials. The gap between those two figures is the state of medical AI, and no vendor page will walk you through it.
What the list actually is
The FDA maintains a public, periodically updated list of AI-enabled medical devices that have been authorized for marketing in the United States. It is not a comprehensive safety database and it is not current. The September 4, 2026 update listed devices whose latest decision was dated June 29, 2026, a two-month lag that recurs in every release (Complizen analysis of the FDA list). The list counts authorizations, not products, and a large and growing share of authorizations are updated versions of devices already on the list: roughly 34% over 2022 to 2024, up from 14% in 2017 to 2019 (npj Digital Medicine).
Growth is real even with those caveats. Annual counts on the list rose from around 65 devices in 2018 to the low hundreds, with 2025 the biggest year recorded (between 295 and 335 depending on the snapshot and counting method; Innolitics counted 295 clearances from 221 unique manufacturers). The 2026 pace is tracking slightly ahead of 2025: 181 authorizations through June 29 against 173 for the same period a year earlier.
The pathway is the story
About 96% of these clearances arrive through the 510(k) pathway, according to a third-party analysis of the FDA's open database (Complizen). A 510(k) does not require proof that a device improves patient outcomes. It requires a demonstration of "substantial equivalence" to a predicate device that was itself cleared earlier, often years earlier, often without outcome data. Researchers who mapped these predicate networks have warned that evidence gaps propagate down the chain rather than getting closed (The Lancet Digital Health, via the PLOS census).
This is not a loophole. It is the design of the pathway, which handles most moderate-risk devices, from powered wheelchairs to ultrasound systems. Applied to software, it means an FDA clearance for an AI radiology tool certifies market access, not clinical effectiveness. A new algorithm can be cleared by showing it performs comparably to an old algorithm on benchmark images. No trial, no patient endpoint, no long-term follow-up required.
What is inside the box
The registry is far narrower than the phrase "medical AI" suggests. The Stanford-led taxonomy of the first 1,016 authorizations found that 84.4% of the devices take images as the AI input, and radiology alone is the review panel for the overwhelming majority of those. The dominant AI function is not diagnosis but quantification and feature localization: measuring structures, segmenting anatomy, flagging regions. Among devices that generate data rather than analyze it, 79% do image enhancement, meaning denoising and accelerated reconstruction (npj Digital Medicine). Radiology accounts for 71.5% of 2025 clearances by Innolitics' count and 78% of all cleared devices by the PLOS census. The single most common product code, QIH, "Radiological Computer Assisted Detection/Diagnosis," covered 75 clearances in 2025 alone.
The same taxonomy review found no evidence of large language models anywhere in the device list as of late 2024. The chat assistants, clinical scribes and reasoning agents dominating health AI headlines live outside this registry almost entirely, because most are not marketed as diagnostic devices. When LLM-based tools do seek clearance, reviewers note, they may not fit existing taxonomy classes at all.

The manufacturer picture is also different from the startup press releases. Of 295 clearances in 2025, 183 went to companies with exactly one. Only nine companies got four or more. The single largest clearing manufacturer was Shanghai United Imaging Healthcare, one of the imaging hardware giants; more than half of the AI 510(k) holders sit outside the United States, with South Korea among the most active countries (Complizen openFDA analysis). Much of the list is established scanner and monitor vendors bolting a software badge onto existing products, not an AI-native industry reshaping medicine.
The evidence census
The structural problem with the registry is invisible unless you link it to trial databases, which is exactly what a PLOS Digital Health study published in August 2026 did. Of 1,357 cleared AI devices as of December 2025:
- 34 (2.5%) were linked to a registered prospective trial
- 12 (0.9%) had trial results posted
- 12 (0.9%) had peer-reviewed publications
- 3 (0.2%) evaluated patient-centered outcomes such as mortality, major morbidity or readmissions
Radiology, with 1,059 cleared devices, had three linked trials. Anesthesiology, with 22 devices, had none. Of the trials that do exist, 94% are industry-sponsored, 82% use diagnostic accuracy or other surrogate endpoints, and two thirds were conducted only in the United States on small, homogeneous cohorts.
The consequences are not hypothetical. The authors point to the Epic Sepsis Model, deployed in more than half of US hospitals, which failed to flag 67% of patients who developed sepsis while alerting on 18% of all hospitalized patients. They cite IBM Watson for Oncology, which reached only 33% concordance with clinicians in Denmark and 12% for gastric cancer in China despite its FDA clearance. A separate analysis of 950 FDA-cleared AI devices found 6% associated with recall events, nearly half within the first year of clearance.

Even the escape hatch is underused
The FDA anticipated the mismatch between static clearances and continuously updated models. Its Predetermined Change Control Plan (PCCP) guidance lets a manufacturer pre-authorize a bounded roadmap of algorithm updates, so improvements do not require a fresh submission every quarter. Uptake has been thin: 30 of 295 clearances in 2025, about 10%, included a PCCP (Innolitics). The industry that lobbies hardest for faster AI review has mostly not used the mechanism that would let it iterate after clearance without re-filing.
The 2026 pace shows the category slowly diversifying. Radiology's share of monthly clearances dropped from around 79% to 56% in April 2026; TytoCare received a De Novo classification for the first AI-powered otoscope after a 330-day review, opening an ENT lane that had five prior AI clearances; digital pathology now counts around 51 authorized devices (Innolitics monthly briefs). De Novo matters because, unlike 510(k), it requires an independent demonstration of safety and effectiveness for novel devices. It remains rare.
What the numbers mean for anyone evaluating medical AI
For readers covering AI, the FDA registry is the closest thing medical AI has to a leaderboard, and it measures the wrong race. A clearance count tells you a company navigated an equivalence submission. It does not tell you the tool improves outcomes, because the file does not need to contain that evidence, and in 99.8% of cases no such evidence exists anywhere.
The useful questions when a vendor says "FDA-cleared AI": which pathway (510(k), De Novo, PMA); which predicate, and what clinical evidence that predicate itself has; and whether any prospective study with a patient endpoint exists, registered and published, not a marketing whitepaper. On current data, the honest base rate answer to the last question is three in 1,357.
Regulatory approval has outpaced clinical validation, and the gap is now the whole story. The clearances are real. The evidence is three.
Sources
- FDA, Artificial Intelligence-Enabled Medical Devices (summary and list, updated September 4, 2026): https://www.fda.gov/medical-devices/digital-health-center-excellence/artificial-intelligence-enabled-medical-devices
- FDA, List of AI-Enabled Medical Devices: https://www.fda.gov/medical-devices/artificial-intelligence-enabled-medical-devices/list-artificial-intelligence-enabled-medical-devices
- Abulibdeh et al., "1,357 AI medical devices cleared, 3 actually tested on patient outcomes," PLOS Digital Health, August 2026: https://pubmed.ncbi.nlm.nih.gov/42616721/
- Singh et al., "How AI is used in FDA-authorized medical devices: a taxonomy across 1,016 authorizations," npj Digital Medicine, 2025: https://www.nature.com/articles/s41746-025-01800-1
- Innolitics, "2025 Year in Review: AI/ML Medical Device 510(k) Clearances," December 2025: https://innolitics.com/articles/year-in-review-ai-ml-medical-device-k-clearances
- Innolitics, "April 2026: 26 AI FDA 510(k) and One De Novo," May 2026: https://innolitics.com/articles/april-ai-fda-k-and-one-de-novo/
- Complizen, "FDA Has Cleared 1,614 AI Medical Devices" (third-party openFDA analysis, September 2026 snapshot): https://www.complizen.ai/post/fda-has-cleared-1614-ai-medical-devices